- Contact Person : Mr. Shi Steven
- Company Name : Shanghai Luheng Instrument Co., Ltd.
- Tel : 86-21-24287122
- Fax : 86-21-54792677
- Address : Shanghai,Shanghai,Room 125,Buidling 7,No.318 Shangwan Road,Nanqiao Town,Fengxian District,Shanghai
- Country/Region : China
Open Cup Flash Point Tester
Open Cup Flash Point Tester
Model: SYD-267
DescriptionLH-901 Cup Flash Point Tester is designed and made as per the National Standard GB/T267 “Test Methods for Flash Point and Fire Point of Petroleum Products (Open Cup Methods)” and Industry Standard of Petrochemical SH/T0318 “Technical Requirements for Open Cup Flash Point Tester”. It is used to determine the flash point and fire point of lubricating oil and Dark petroleum products.Technical parameters1. Power supply: AC 220 V±10%,50Hz;2. Outer crucible:1) Material: Carbon structural steel;2) Diameter of upper opening: Φ100±5 mm;3) Height: 50±5 mm;4) Diameter of bottom: Φ56±2 mm;3. Inner crucible:1) Material: Carbon structural steel;2) Diameter of upper opening: Φ64±1 mm;3) Height: 47±1 mm;4) Diameter of bottom: Φ38±1 mm;5) Scale line: there is a line 12 mm away from the rim of upper opening and a line 18 mm away from the rim of upper opening.4. Gas tube: the diameter of nozzle isΦ0.8~Φ1 mm. The inner surface of tube is smooth and it can adjust flame to 3 to 4 mm.5. Heater: Heated by an electric furnace. Heating power is 1000 W and adjustable.6. Thermometer: 0~360 °C; each scale division is 1 °C. It conforms to GB/T514.7. Ambient temperature: ≤35 °C;8. Relative humidity: ≤85%;9. Total power consumption: not more than 1100W.
Open Cup Flash Point Tester,Open cup point apparatus
Open Cup Flash Point Tester