Cleveland Flash Point Apparatus

Cleveland Flash Point Apparatus
Product Detailed
Related Categories:Other Analysis Instruments
Cleveland Flash Point Apparatus It is suitable to determine flash point 1.AC 220V, 50 Hz 2.Not more than 650 W

 

Cleveland Flash Point Apparatus

 

The instrument is designed and made as per the National Standard of People’s Republic of China GB/T3536-2008 “Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Method)”. It is suitable to determine flash point and fire point of petroleum products and asphalt, excepting fuel oils and petroleum products having open cup flash point lower than 79 °C.

 

I. Main technical characteristics

1. It adopts a special heating furnace for operation safety. The heating power can be continuously adjusted for the requirement of test.

2. The instrument adopts desktop structure. It is with stainless steel workbench and stainless steel shell of heating furnace, so it is beautiful.

3. It is equipped with complete accessories. You only need to prepare coal gas or other civil combustible gas before test.

4. It is equipped with a fire extinguishing cover. It is in accord with the requirement of fire point test.

 

II. Main technical specification and parameters

1. Power supply: AC 220V±10%, 50 Hz;

2. Heating device: quartz tube electric furnace; there is no flame and it is anti-explosive. The heating power can be continuously adjusted in the range of 0∼600W;

3. Thermometer: -6~400°C; Scale division is 2 °C; as per GB/T514;

4. Flame application device:

(1) Gas source: coal gas (or other civil combustible gas)

(2) Spout diameter: 0.6∼0.8 mm

5. Ambient temperature: -10°C∼50°C;

6. Relative humidity: ≤85%;

 

 

 

Cleveland Flash Point Apparatus